Precision Medical PM5950 AccuO2 Oxygen Analyzer
The AccuO2 Oxygen Analyzer is designed to analyze and measure a devices oxygen purity, outlet pressure and flow accuracy. These features allow users to save on costs and time when checking a patient’s device. The analyzer is a compact handheld device is easy to transport, operate and store for the user’s convenience. The LCD screen display is easy to read, with fast results in a large, bold font that allows for a quicker setup time.
Features:
- Lightweight
- Handheld or mountable
- Easy one touch calibration using medical air or oxygen
- 5,000+ hour operation using 2 AA batteries
Specifications:
Base Device Specifications
Dimensions: (Analyzer without Cable and Sensor attached):
Depth: 1.30” (3.32 cm)
Width: 2.90” (7.30 cm)
Height: 4.00” (10.23 cm)
Cable Length: 10 ft. (3.05m) (fully extended)
Weight:
Device Weight: 0.35 lbs (5.60 oz / 0.15 kg)
(includes: Analyzer, Sensor, Cable and batteries)
CAUTION DO NOT:
use if dirt or contaminants are present on or around this Oxygen Analyzer or connecting devices.
smoke in an area where oxygen is being administered.
clean with aromatic hydrocarbons.
steam autoclave.
gas sterilize.
immerse Oxygen Analyzer or Sensor in liquid.
immerse the sensor in any cleaning solution, autoclave or
expose the sensor to high temperatures.
The Oxygen Sensor is a sealed device containing a mild acid
electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are
hazardous waste and should be disposed of properly.
Dropping Sensor can adversely affect its performance.
Be sure to apply 100% oxygen, or ambient air concentration to the
device during calibration or the device will not calibrate correctly.
Temperature: 50ºF – 113ºF (10ºC – 45ºC)
Altitude: Sea Level to 8000 feet
Humidity: 0 – 95% non-condensing
Storage Conditions
Temperature: 5°F – 122°F (-15°C – 50°C)
Humidity: 0 – 95% non-condensing
Mode of Operation: Continuous
Electrical Classification: Internally powered Medical Electrical equipment
Diverter Fitting: fits industry standard, 15 mm Tee Adapter
Measurement Range: 0.0 – 100% Oxygen
Resolution: 0.1 %
Total Accuracy: ± 3.0% Actual Oxygen Level over full operating temperature range
Drift of Measurement: < +/-1% of full scale at constant temperature, pressure and humidity
Response Time: 90% of final value in less than 12 seconds at 77°F (25ºC)
Warm-up Time: Not required
Low Battery: Low battery icon displayed
Indication: Indirect contact via gas passing through
Patient Contact: sensor sampling site.
Sensor Specifications
Sensor Type: Galvanic Oxygen Sensor; Precision Medical PN 504877
Expected Sensor Life: > 1,000,000 O2 % Hours