Amico M-PRSW-GA-ORES Medical Gas Pressure Switch For Reserve Manifold
The Reserve Manifold Pressure Switch is used for monitoring the supply pressure in the reserve system. The switch is preset @ 1,100 psi [7,585 kPa] decreasing pressure, to alarm before the reserve header contents fall to one day’s average supply, indicating reserve low as per NFPA and CSA.
Electrical Characteristics: The electrical rating (continuous inductive) 10 amps 125 or 250 volts AC, 3 amps 480 volts AC. AutomatIcally reset by snap-action of switch.
Electrical Connection: Normally open/normally closed contacts. Electrical connections are made through a 1⁄2” conduit connection to screw terminals on the switch. Pressure connection: 1⁄4” NPT.
Adjustment Instructions: Turn adjustment screw clockwise to lower actuation point (switch setting).
The switch enclosure shall be a Nema 4 water tight housing with mounting bracket, 1/4”FNPT bottom inlet gas service line connection and a 1/2” conduit connection on the top.
• Instrument quality snap-action switch
• Normally open or normally closed contacts
• Nema 4 water-tight housing with mounting brackets • Pressure switches for normal reserve systems
• Proof pressure 4,800 psi [33,100 kPa]
• Adjustable range: 160-3,200 psi [1,105-22,060 kPa]
• Preset @ 1,100 psi [7,585 kPa] decreasing
• Factory cleaned and sealed for oxygen service